An automatic immuno-microfluidic system integrating electrospun polystyrene microfibrous reactors for rapid detection of salivary cortisol

集成电纺聚苯乙烯微纤维反应器的自动免疫微流体系统用于快速检测唾液皮质醇

阅读:10
作者:Yecan Wang, Hiroshi Murakami, Toshihiro Kasama, Shigenobu Mitsuzawa, Satoru Shinkawa, Ryo Miyake, Madoka Takai

Abstract

Conventional competitive enzyme-linked immunosorbent assay (ELISA) to measure the cortisol level in body fluid consumes a large amount of time, owing to complicated operations involved and requirement for precise control of reagent addition. We developed an automatic microfluidic system to detect salivary cortisol rapidly, and an electrospun polystyrene (PS) microfiber-based reactor providing considerable binding sites for antibody immobilization, thus resolving the time limitations of competitive ELISA. Cortisol sample, horseradish peroxidase (HRP)-conjugated cortisol, and 3,3',5,5'-tetramethylbenzidine (TMB) substrate were delivered to the PS reactor from containers in sequence by pumps automatically. The color variation due to oxidized TMB complex reflects the cortisol concentration level measured using an RGB phototransistor. In addition, the entire procedure from sample introduction to obtaining the photocurrent took only 15 min. This system can be implemented to quantify cortisol from 0.37 ng/mL to 30 ng/mL, and the limit of detection was estimated at 0.37 ng/mL.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。