Following revised information pertaining to newer swab types and testing protocols in the new CLSI M40-A2 standard, we evaluated three liquid swab transport systems for the recovery of aerobic, anaerobic, and fastidious organisms at room temperature and at 4°C. All tested liquid swab transport systems were fully compliant with the M40-A2 standard, with acceptable performance at both temperatures after the full specified holding period, using both qualitative (roll-plate) and quantitative (swab elution) methods.
Evaluation of Liquid-Based Swab Transport Systems against the New Approved CLSI M40-A2 Standard.
阅读:11
作者:Gizzie Nina, Adukwu Emmanuel
| 期刊: | Journal of Clinical Microbiology | 影响因子: | 5.400 |
| 时间: | 2016 | 起止号: | 2016 Apr;54(4):1152-6 |
| doi: | 10.1128/JCM.03337-15 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。