A PrinCE autosampler was coupled to a Q-Exactive mass spectrometer by an electrokinetically pumped sheath-flow nanospray interface to perform automated capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry (CZE-ESI-MS/MS). 20ng aliquots of an Escherichia coli digest were injected to evaluate the system. Eight sequential injections over an 8-h period identified 1115±70 (relative standard deviation, RSD=6%) peptides and 270±8 (RSD=3%) proteins per run. The average RSDs of migration time, peak intensity, and peak area were 3%, 24% and 19%, respectively, for 340 peptides with high intensity. This is the first report of an automated CZE-ESI-MS/MS system using the electrokinetically pumped sheath-flow nanospray interface. The results demonstrate that this system is capable of reproducibly identifying over 1000 peptides from an E. coli tryptic digest in a 1-h analysis time.
Stable, reproducible, and automated capillary zone electrophoresis-tandem mass spectrometry system with an electrokinetically pumped sheath-flow nanospray interface.
阅读:4
作者:Zhu Guijie, Sun Liangliang, Yan Xiaojing, Dovichi Norman J
| 期刊: | Analytica Chimica Acta | 影响因子: | 6.000 |
| 时间: | 2014 | 起止号: | 2014 Jan 31; 810:94-8 |
| doi: | 10.1016/j.aca.2013.11.057 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。
