Stable, reproducible, and automated capillary zone electrophoresis-tandem mass spectrometry system with an electrokinetically pumped sheath-flow nanospray interface.

阅读:4
作者:Zhu Guijie, Sun Liangliang, Yan Xiaojing, Dovichi Norman J
A PrinCE autosampler was coupled to a Q-Exactive mass spectrometer by an electrokinetically pumped sheath-flow nanospray interface to perform automated capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry (CZE-ESI-MS/MS). 20ng aliquots of an Escherichia coli digest were injected to evaluate the system. Eight sequential injections over an 8-h period identified 1115±70 (relative standard deviation, RSD=6%) peptides and 270±8 (RSD=3%) proteins per run. The average RSDs of migration time, peak intensity, and peak area were 3%, 24% and 19%, respectively, for 340 peptides with high intensity. This is the first report of an automated CZE-ESI-MS/MS system using the electrokinetically pumped sheath-flow nanospray interface. The results demonstrate that this system is capable of reproducibly identifying over 1000 peptides from an E. coli tryptic digest in a 1-h analysis time.