Aggregates cause a fatal problem in the structural analysis of a biomacro-mol-ecule in solution using small-angle X-ray or neutron scattering (SAS): they deteriorate the scattering profile of the target molecule and lead to an incorrect structure. Recently, an integrated method of analytical ultracentrifugation (AUC) and SAS, abbreviated AUC-SAS, was developed as a new approach to overcome this problem. However, the original version of AUC-SAS does not offer a correct scattering profile of the target molecule when the weight fraction of aggregates is higher than ca 10%. In this study, the obstacle point in the original AUC-SAS approach is identified. The improved AUC-SAS method is then applicable to a solution with a relatively larger weight fraction of aggregates (â¤20%).
Derivation of the small-angle scattering profile of a target biomacromolecule from a profile deteriorated by aggregates. AUC-SAS.
阅读:3
作者:Morishima Ken, Inoue Rintaro, Sugiyama Masaaki
| 期刊: | Journal of Applied Crystallography | 影响因子: | 2.800 |
| 时间: | 2023 | 起止号: | 2023 Apr 25; 56(Pt 3):624-632 |
| doi: | 10.1107/S1600576723002406 | ||