Use of the labscale tangential flow diafiltrator to remove tetrameric hemoglobin from polyhemoglobin, purify hemolysate, and concentrate polyhemoglobin.

阅读:4
作者:Stefanescu Ada, Chang Thomas Ming Swi
The Labscale tangential flow diafiltration system in conjunction with polyethersulfone Biomax filters (both purchased from Millipore Inc.) were used in the following studies: (1) A crucial step in the preparation of polyhemoglobin (polyHb) is the removal of tetrameric hemoglobin (Hb), which can cause adverse side-effects. The efficiency for this depends on the integrity of the 100kDa filters. Those with lower integrity were not effective whereas those with a higher integrity of 0.075ml/min were more effective. A filter with an integrity of 0.075ml/min can reduce the initial 11.2% tetrameric Hb concentration to 1.3% using a flow rate of 0.2ml/min. Higher flow rate of 2.2ml/min was not as effective. (2) Filtration with a 100kDa filter was quick and efficient in separating stroma-free hemoglobin solutions from cellular debris from hemolysate. (3) Both 5kDa and 100kDa filters are efficient in concentrating hemoglobin solutions.