Post-operative MRI of patients with deep brain simulation (DBS) implants is useful to assess complications and diagnose comorbidities, however more than one third of medical centers do not perform MRIs on this patient population due to stringent safety restrictions and liability risks. A new system of reconfigurable magnetic resonance imaging head coil composed of a rotatable linearly-polarized birdcage transmitter and a close-fitting 32-channel receive array is presented for low-SAR imaging of patients with DBS implants. The novel system works by generating a region with low electric field magnitude and steering it to coincide with the DBS lead trajectory. We demonstrate that the new coil system substantially reduces the SAR amplification around DBS electrodes compared to commercially available circularly polarized coils in a cohort of 9 patient-derived realistic DBS lead trajectories. We also show that the optimal coil configuration can be reliably identified from the image artifact on B(1)(+) field maps. Our preliminary results suggest that such a system may provide a viable solution for high-resolution imaging of DBS patients in the future. More data is needed to quantify safety limits and recommend imaging protocols before the novel coil system can be used on patients with DBS implants.
Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants.
阅读:4
作者:Golestanirad Laleh, Iacono Maria Ida, Keil Boris, Angelone Leonardo M, Bonmassar Giorgio, Fox Michael D, Herrington Todd, Adalsteinsson Elfar, LaPierre Cristen, Mareyam Azma, Wald Lawrence L
| 期刊: | Neuroimage | 影响因子: | 4.500 |
| 时间: | 2017 | 起止号: | 2017 Feb 15; 147:577-588 |
| doi: | 10.1016/j.neuroimage.2016.12.056 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。