Data-science-guided calibration curve prediction of an MLCT-based ee determination assay for chiral amines.

阅读:7
作者:Howard James R, Shuluk Julia R, Bhakare Arya, Anslyn Eric V
Circular dichroism (CD) based enantiomeric excess (ee) determination assays are optical alternatives to chromatographic ee determination in high-throughput screening (HTS) applications. However, the implementation of these assays requires calibration experiments using enantioenriched materials. We present a data-driven approach that circumvents the need for chiral resolution and calibration experiments for an octahedral Fe(II) complex (1) used for the ee determination of α-chiral primary amines. By computationally parameterizing the imine ligands formed in the assay conditions, a model of the circular dichroism (CD) response of the Fe(II) assembly was developed. Using this model, calibration curves were generated for four analytes and compared to experimentally generated curves. In a single-blind ee determination study, the ee values of unknown samples were determined within 9% mean absolute error, which rivals the error using experimentally generated calibration curves.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。