This publication describes the development of an automated platform for the study of the plasma glycoproteome. The method consists of targeted depletion in-line with glycoprotein fractionation. A key element of this platform is the enabling of high throughput sample processing in a manner that minimizes analytical bias in a clinical sample set. The system, named High Performance Multi-Lectin Affinity Chromatography (HP-MLAC), is composed of a serial configuration of depletion columns containing anti-albumin antibody and protein A with in-line multilectin affinity chromatography (M-LAC) which consists of three mixtures of lectins concanavalin A (ConA), jacalin (JAC), and wheat germ agglutinin (WGA). We have demonstrated that this platform gives high recoveries for the fractionation of the plasma proteome (> or = 95%) and excellent stability (over 200 runs). In addition, glycoproteomes isolated using the HP-MLAC platform were shown to be highly reproducible and glycan specific as demonstrated by rechromatography of selected fractions and proteomic analysis of the unbound (glycoproteome 1) and bound (glycoproteome 2) fractions.
Automated platform for fractionation of human plasma glycoproteome in clinical proteomics.
阅读:4
作者:Kullolli Majlinda, Hancock William S, Hincapie Marina
| 期刊: | Analytical Chemistry | 影响因子: | 6.700 |
| 时间: | 2010 | 起止号: | 2010 Jan 1; 82(1):115-20 |