Practical Guide to the Design, Fabrication, and Calibration of NIST Nanocalorimeters

NIST纳米量热仪的设计、制造和校准实用指南

阅读:1

Abstract

We report here on the design, fabrication, and calibration of nanocalorimeter sensors used in the National Institute of Standards and Technology (NIST) Nanocalorimetry Measurements Project. These small-scale thermal analysis instruments are produced using silicon microfabrication approaches. A single platinum line serves as both the heater and temperature sensor, and it is made from a 500 μm wide, 100 nm thick platinum trace, suspended on a 100 nm thick silicon nitride membrane for thermal isolation. Supplemental materials to this article (available online) include drawing files and LabVIEW code used in the fabrication and calibration process.