Comparison of the NIST and PTB Air-Kerma Standards for Low-Energy X-Rays

NIST 和 PTB 低能 X 射线空气比释动能标准的比较

阅读:1

Abstract

A comparison has been made of the air-kerma standards for low-energy x rays at the National Institute of Standards and Technology (NIST) and the Physikalisch-Technische Bundesanstalt (PTB). The comparison involved a series of measurements at the PTB and the NIST using the air-kerma standards and two NIST reference-class transfer ionization chamber standards. Results are presented for the reference radiation beam qualities in the range from 25 kV to 50 kV for low energy x rays, including the techniques used for mammography dose traceability. The tungsten generated reference radiation qualities, between 25 kV and 50 kV used for this comparison, are new to NIST; therefore this comparison will serve as the preliminary comparison for NIST and a verification of the primary standard correction factors. The mammography comparison will repeat two previously unpublished comparisons between PTB and NIST. The results show the standards to be in reasonable agreement within the standard uncertainty of the comparison of about 0.4 %.