Automated method for the determination of the band gap energy of pure and mixed powder samples using diffuse reflectance spectroscopy

使用漫反射光谱法自动测定纯粉末和混合粉末样品的带隙能量

阅读:9

Abstract

An automated method to determine the band gap energy (E g ) of pure and mixed powder compounds using diffuse reflectance spectroscopy is presented. This method is based on a five-step algorithm that mimics the judgment made by an expert analyst in identifying the linear segments in Tauc plots and subsequent estimation of the E g value. It is demonstrated that the method to estimate E g by intersecting the straight-line fit of the Tauc segment with the photon energy axis is not appropriate for those samples containing more than one optical absorbing phase because systematic underestimation of the E g value results. The automated method accounts for such cases by introducing a base line function. The robustness of the implemented algorithm was tested using three model systems, ZnO-Al2O3, ZnO-CoO and ZnO-CdO. The estimated E g 's using the automated method differ in less than 1% than those obtained by its manual counterpart.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。