Comparative study of different SARS-CoV-2 diagnostic techniques

不同SARS-CoV-2诊断技术的比较研究

阅读:2

Abstract

The rapid spread of SARS-CoV-2 led to the necessity of developing diagnostic tests for rapid virus detection. Many commercial platforms have appeared and have been approved for this purpose. In this study, 95 positive and 5 negative retrospective samples were analyzed by 4 different commercial RT-qPCR kits (TaqMan 2019nCoV Assay, Allplex™SARS-COV-2 Assay, FTD SARS-COV-2 Assay and qCOVID-19). The Hologic Aptima SARS-COV-2 and the Clart-COVID-19 system were also tested. serial dilutions of SARS-COV-2 standard control were included for sensitivity analysis. Among the qPCR tested qCOVID19 and Allplex™SARS-COV-2 Assay were both able to detect all the clinical samples included in the study. All four qPCR evaluated showed high sensitivity for samples with Ct<33. Clart-COVID-19 microarrays detected all samples and controls used in this study whereas Hologic Aptima Panther failed with one of the clinical samples. However, the main problem with this system was the number of invalidated samples despite avoiding the use of medium with guanidine isothiocyanate as recommended by the manufacturer. All the techniques tested were of value for SARS-CoV-2 detection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。