Variable classifications of glycemic index determined by glucose meters

血糖仪测定的血糖指数的多种分类

阅读:1

Abstract

THE STUDY EVALUATED AND COMPARED THE DIFFERENCES OF GLUCOSE RESPONSES, INCREMENTAL AREA UNDER CURVE (IAUC), GLYCEMIC INDEX (GI) AND THE CLASSIFICATION OF GI VALUES BETWEEN MEASURED BY BIOCHEMICAL ANALYZER (FUJI AUTOMATIC BIOCHEMISTRY ANALYZER (FAA)) AND THREE GLUCOSE METERS: Accue Chek Advantage (AGM), BREEZE 2 (BGM), and Optimum Xceed (OGM). Ten healthy subjects were recruited for the study. The results showed OGM yield highest postprandial glucose responses of 119.6 +/- 1.5, followed by FAA, 118.4 +/- 1.2, BGM, 117.4 +/- 1.4 and AGM, 112.6 +/- 1.3 mg/dl respectively. FAA reached highest mean IAUC of 4156 +/- 208 mg x min/dl, followed by OGM (3835 +/- 270 mg x min/dl), BGM (3730 +/- 241 mg x min/dl) and AGM (3394 +/- 253 mg x min/dl). Among four methods, OGM produced highest mean GI value than FAA (87 +/- 5) than FAA, followed by BGM and AGM (77 +/- 1, 68 +/- 4 and 63 +/- 5, p<0.05). The results suggested that the AGM, BGM and OGM are more variable methods to determine IAUC, GI and rank GI value of food than FAA. The present result does not necessarily apply to other glucose meters. The performance of glucose meter to determine GI value of food should be evaluated and calibrated before use.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。