Influence of fixation conditions on the performance of glutaraldehyde-treated porcine aortic valves: towards a more scientific basis

固定条件对戊二醛处理猪主动脉瓣膜性能的影响:迈向更科学的基础

阅读:1

Abstract

To maintain optimum mechanical properties in glutaraldehyde-treated heart-valve tissue the full collagen crimp geometry originally present in the relaxed fresh tissue should be retained. By varying the pressure at which glutaraldehyde fixation is carried out, considerable alterations to this crimp geometry can be achieved. The mechanical stiffness of the preserved tissue is consequently affected, and this in turn has a striking influence on both the opening behaviour of the valve and the degree of strain localisation in the leaflet tissue. A pressure of 100 mmHg eliminated the collagen crimp geometry entirely, and this resulted in the formation of sites of local strain or kinks in the valve leaflets during opening. It is expected that this strain localisation phenomenon will influence the long-term fatigue durability of the treated tissue. Pressures even as low as 4 mmHg result in significant reductions of crimp geometry. Fresh valves should therefore be fixed under a positive head of pressure sufficient only to ensure that the leaflets seal along their coapting free margins. A pressure of less than 1 mmHg was sufficient to achieve this. Leaflets of the commercially available Hancock valve show features similar to valves fixed in glutaraldehyde at about 100 mmHg pressure.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。