Feasibility of the Lilium α-200 portable ultrasound bladder scanner for accurate bladder volume measurement

Lilium α-200便携式超声膀胱扫描仪用于精确测量膀胱容量的可行性

阅读:1

Abstract

PURPOSE: The aim of this study was to investigate whether data obtained from the Lilium α-200 (Lilium Otsuka Co., Ltd., Japan) correlated with conventional frequency-volume chart (FVC) and post-void residual urine volume (PVR) obtained by urethral catheterization. MATERIALS AND METHODS: This was a prospective multicentre study. Patients hospitalized for the treatment of lower urinary tract symptoms were included. Patients were evaluated with conventional FVC and Lilium α-200 for 2 days. PVR was measured by urethral catherization after urination at the end of the 2 day evaluation period. RESULTS: A total of 42 patients were enrolled in this study. Voided volume and PVR measured by Lilium α-200 were significantly correlated with voided volume obtained from conventional FVC and PVR measured by urethral catheterization, respectively. There was considerable measurement error in voided volume measured by Lilium α-200 (-21.0±102.0 mL). In contrast, the error between PVR measured by the Lilium α-200 and PVR obtained by urethral catheterization was 2.4±52.0 mL. Additionally, high body mass index, but not sex, benign prostate hyperplasia, time zone of measurement (daytime vs. nighttime), and examiners (a urologist versus other healthcare providers) were significantly associated with inaccurate results in voided volume. CONCLUSIONS: Voided volume and PVR measured by the Lilium α-200 were correlated with voided volume obtained from conventional FVC and PVR measured by urethral catheterization, although accuracy of the measurements was not high. The Lilium α-200 is a useful device to easily measure approximate bladder volume.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。