Numerical simulation of in vivo intraosseous torsional failure of a hollow-screw oral implant

空心螺钉口腔种植体体内骨内扭转失效的数值模拟

阅读:1

Abstract

BACKGROUND: Owing to the complexity and magnitude of functional forces transferred to the bone-implant interface, the mechanical strength of the interface is of great importance. The purpose of this study was to determine the intraosseous torsional shear strength of an osseointegrated oral implant using 3-D finite element (FE) stress analysis implemented by in vivo failure torque data of an implant. METHODS: A Ø 3.5 mm x 12 mm ITI hollow screw dental implant in a patient was subjected to torque failure test using a custom-made strain-gauged manual torque wrench connected to a data acquisition system. The 3-D FE model of the implant and peri-implant circumstances was constructed. The in vivo strain data was converted to torque units (N x cm) to involve in loading definition of FE analysis. Upon processing of the FE analysis, the shear stress of peri-implant bone was evaluated to assume torsional shear stress strength of the bone-implant interface. RESULTS: The in vivo torque failure test yielded 5952 microstrains at custom-made manual torque wrench level and conversion of the strain data resulted in 750 N.cm. FE revealed that highest shear stress value in the trabecular bone, 121 MPa, was located at the first intimate contact with implant. Trabecular bone in contact with external surface of hollow implant body participated shear stress distribution, but not the bone resting inside of the hollow. CONCLUSION: The torsional strength of hollow-screw implants is basically provided by the marginal bone and the hollow part has negligible effect on interfacial shear strength.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。