Noninvasive Blood Pressure Monitor Designed for Patients With Heart Failure Supported with Continuous-Flow Left Ventricular Assist Devices

专为接受连续流左心室辅助装置治疗的心力衰竭患者设计的无创血压监测仪

阅读:1

Abstract

The gold standard for noninvasive blood pressure (BP) measurement, the Doppler technique, does not provide systolic blood pressure (SBP) and diastolic blood pressure (DBP) and may limit therapy outcomes. To improve patient care, we tested specifically designed experimental BP (ExpBP) monitor and the Doppler technique by comparing noninvasive measures to the intraarterial (I-A) BP in 31 patients with end-stage heart failure (4 females) 2.6 ± 3.4 days post-LVAD implantation (20 HeartMate II and 11 HeartWare). Bland-Altman plots revealed that the ExpBP monitor overestimated mean arterial pressure (MAP) by 1.2 (4.8) mm Hg (mean difference [standard deviation]), whereas the Doppler by 6.7 (5.8) mm Hg. The ExpBP SBP was overestimated by 0.8 (6.1) mm Hg and DBP by 1.9 (5.3) mm Hg compared with the respective I-A pressures. Both techniques achieved similar measurement reliability. In the measurement "success rate" expressed as a frequency (percent) of readable BP values per measurement attempts, Doppler accomplished 100% vs. 97%, 97%, and 94% of successful detections of MAP, SBP, and DBP provided by the ExpBP monitor. The ExpBP monitor demonstrated higher accuracy in the MAP assessment than the Doppler in addition to providing SBP and DBP in majority of subjects. Improved BP control may help to mitigate related neurologic adverse event rates.