Intermittent Aortic Valve Opening and Risk of Thrombosis in Ventricular Assist Device Patients

间歇性主动脉瓣开放与心室辅助装置患者血栓形成风险

阅读:1

Abstract

The current study evaluates quantitatively the impact that intermittent aortic valve (AV) opening has on the thrombogenicity in the aortic arch region for patients under left ventricular assist device (LVAD) therapy. The influence of flow through the AV, opening once every five cardiac cycles, on the flow patterns in the ascending aortic is measured in a patient-derived computed tomography image-based model, after LVAD implantation. The mechanical environment of flowing platelets is investigated, by statistical treatment of outliers in Lagrangian particle tracking, and thrombogenesis metrics (platelet residence times and activation state characterized by shear stress accumulation) are compared for the cases of closed AV versus intermittent AV opening. All hemodynamics metrics are improved by AV opening, even at a reduced frequency and flow rate. Residence times of platelets or microthrombi are reduced significantly by transvalvular flow, as are the shear stress history experienced and the shear stress magnitude and gradients on the aortic root endothelium. The findings of this device-neutral study support the multiple advantages of management that enables AV opening, providing a rationale for establishing this as a standard in long-term treatment and care for advanced heart failure patients.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。