Evaluation of local gas exchange in a pulsating respiratory support catheter

脉动式呼吸支持导管局部气体交换的评估

阅读:1

Abstract

An intravenous respiratory support catheter, the next generation of artificial lungs, is being developed in our laboratory to potentially support acute respiratory failure or patients with chronic obstructive pulmonary disease with acute exacerbations. A rapidly pulsating 25 ml balloon inside a bundle of hollow fiber membranes facilitates supplemental oxygenation and CO2 removal. In this study, we hypothesized that non-uniform gas exchange in different regions of this fiber bundle was present because of asymmetric balloon collapse and the interaction of longitudinal flow. Four quarter regions and two rings around the central balloon were selectively perfused to evaluate local gas exchange in a 3.18 cm test section using helium as the sweep gas. Quarter region CO2 exchange rates at 400 beats per minute were 156.8 +/- 0.8, 162.5 +/- 1.8, 157.2 +/- 0.2, and 196.6 +/- 0.8 ml/min/m2 (top, front, bottom, and back, respectively). The back section, adjacent to convex balloon collapse, had 17-20% higher exchange than the other sections caused by higher relative velocities past its stationary fibers. Inner and outer ring maximum pulsation gas exchange rates were 174.4 +/- 1.8 and 174.6 +/- 0.9 ml/min/m2, respectively, showing that fluid flow was equally distributed throughout the fiber bundle.