Pulmonary Artery Perforation by Atrial Appendage Closure Device and the Contribution of Auricular Contraction to Its Pathogenesis

心耳封堵器引起的肺动脉穿孔及其发病机制中心耳收缩的作用

阅读:1

Abstract

We report the case of an 83-year-old man with paroxysmal atrial fibrillation who underwent successful percutaneous left atrial appendage closure with the LAmbre device, being in sinus rhythm at implantation. Suddenly, the patient experienced cardiac tamponade and died within a few minutes. Autopsy revealed a slight protrusion of the LAmbre device into the atrial appendage wall, and pulmonary artery laceration. This is the first published report of pulmonary artery perforation by the LAmbre device. This case highlights the need for a detailed imaging study before this procedure is performed, to assess left atrial appendage movement/contraction in patients in sinus rhythm.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。