Application of a Waveguide-Mode Sensor to Blood Testing for Hepatitis B Virus, Hepatitis C Virus, Human Immunodeficiency Virus and Treponema pallidum Infection

将波导模式传感器应用于乙型肝炎病毒、丙型肝炎病毒、人类免疫缺陷病毒和梅毒螺旋体感染的血液检测

阅读:1

Abstract

Testing for blood-transmitted infectious agents is an important aspect of safe medical treatment. During emergencies, such as significant earthquakes, many patients need surgical treatment and/or blood transfusion. Because a waveguide mode (WM) sensor can be used as a portable, on-site blood testing device in emergency settings, we have previously developed WM sensors for detection of antibodies against hepatitis B virus and hepatitis C virus and for forward ABO and Rh(D) and reverse ABO blood typing. In this study, we compared signal enhancement methods using secondary antibodies conjugated with peroxidase, a fluorescent dye, and gold nanoparticles, and found that the peroxidase reaction method offers superior sensitivity while gold nanoparticles provide the most rapid detection of anti-HBs antibody. Next, we examined whether we could apply a WM sensor with signal enhancement with peroxidase or gold nanoparticles to detection of antibodies against hepatitis C virus, human immunodeficiency virus and Treponema pallidum, and HBs antigen in plasma. We showed that a WM sensor can detect significant signals of these infectious agents within 30 min. Therefore, a portable device utilizing a WM sensor can be used for on-site blood testing of infectious agents in emergency settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。