Anatomical considerations for the development of a new transcatheter aortopulmonary shunt device in patients with severe pulmonary arterial hypertension

针对重度肺动脉高压患者,开发新型经导管主动脉肺动脉分流装置的解剖学考量

阅读:1

Abstract

Abstract Morbidity from pulmonary arterial hypertension (PAH) ensues when the pulmonary pressure reaches suprasystemic levels. A transcatheter alternative to the Potts shunt would allow decompression of the right heart without the surgical risks. To aid development of a transcatheter aortopulmonary shunt (TAPS) device, we described the anatomic relationship between the left pulmonary artery (LPA) and the descending aorta (dAO) in adults with severe pulmonary hypertension. Adults with severe PAH (peak systolic pulmonary arterial pressure [PASP] ≥80 mmHg) who had computed tomography of the chest were enrolled. Measurements were taken on the axial plane at the level of the pulmonary artery bifurcation. Forty patients (male sex, 9 patients; median age ± standard deviation [SD], [Formula: see text] years; median PASP ± SD, [Formula: see text] mmHg) were identified. The mean distance (±SD) between the LPA and dAO was [Formula: see text] mm. The mean luminal dAO and LPA diameters (±SD) were [Formula: see text] mm and [Formula: see text] mm, respectively. The LPA and dAO approximated in 93% of patients, with 38% having aortic calcification at the contact site. The mean "landing zone" width and height (defining an area with distance <4 mm between the outer borders) of the two arteries were [Formula: see text] mm and [Formula: see text] mm, respectively, at a mean distance of [Formula: see text] mm from the main pulmonary artery bifurcation. This study shows that the landing zone is able to accommodate a TAPS device of up to 15 mm in diameter in the majority of patients with severe PAH.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。