Multicenter US Clinical Trial With an Electric-Acoustic Stimulation (EAS) System in Adults: Final Outcomes

美国多中心临床试验:成人电声刺激(EAS)系统最终结果

阅读:1

Abstract

OBJECTIVE: To demonstrate the safety and effectiveness of the MED-EL Electric-Acoustic Stimulation (EAS) System, for adults with residual low-frequency hearing and severe-to-profound hearing loss in the mid to high frequencies. STUDY DESIGN: Prospective, repeated measures. SETTING: Multicenter, hospital. PATIENTS: Seventy-three subjects implanted with PULSAR or SONATA cochlear implants with FLEX electrode arrays. INTERVENTION: Subjects were fit postoperatively with an audio processor, combining electric stimulation and acoustic amplification. MAIN OUTCOME MEASURES: Unaided thresholds were measured preoperatively and at 3, 6, and 12 months postactivation. Speech perception was assessed at these intervals using City University of New York sentences in noise and consonant-nucleus-consonant words in quiet. Subjective benefit was assessed at these intervals via the Abbreviated Profile of Hearing Aid Benefit and Hearing Device Satisfaction Scale questionnaires. RESULTS: Sixty-seven of 73 subjects (92%) completed outcome measures for all study intervals. Of those 67 subjects, 79% experienced less than a 30 dB HL low-frequency pure-tone average (250-1000 Hz) shift, and 97% were able to use the acoustic unit at 12 months postactivation. In the EAS condition, 94% of subjects performed similarly to or better than their preoperative performance on City University of New York sentences in noise at 12 months postactivation, with 85% demonstrating improvement. Ninety-seven percent of subjects performed similarly or better on consonant-nucleus-consonant words in quiet, with 84% demonstrating improvement. CONCLUSION: The MED-EL EAS System is a safe and effective treatment option for adults with normal hearing to moderate sensorineural hearing loss in the low frequencies and severe-to-profound sensorineural hearing loss in the high frequencies who do not benefit from traditional amplification.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。