Photoacoustic trace gas detection of OCS using a 2.45 mL Helmholtz resonator and a 4823.3 nm ICL light source

利用2.45 mL亥姆霍兹共振器和4823.3 nm ICL光源进行光声痕量气体OCS检测

阅读:1

Abstract

A miniaturized photoacoustic spectroscopy-based gas sensor is proposed for the purpose of detecting sub-ppm-level carbonyl sulfide (OCS) using a tunable mid-infrared interband cascade laser (ICL) and a Helmholtz photoacoustic cell. The tuning characteristics of the tunable ICL with a center wavelength of 4823.3 nm were investigated to achieve the optimal driving parameters. A Helmholtz photoacoustic cell with a volume of ∼2.45 mL was designed and optimized to miniaturize the measurement system. By optimizing the modulation parameters and signal processing, the system was verified to have a good linear response to OCS concentration. With a lock-in amplifier integration time of 10 s, the 1σ noise standard deviation in differential mode was 0.84 mV and a minimum detection limit (MDL) of 409.2 ppbV was achieved at atmospheric pressure and room temperature.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。