Validation testing of five home blood pressure monitoring devices for the upper arm according to the ISO 81060-2:2018/AMD 1:2020 protocol

根据 ISO 81060-2:2018/AMD 1:2020 协议对五种家用上臂血压监测设备进行验证测试

阅读:1

Abstract

The accuracy of Omron 10 Series BP7450 (HEM-7342T-Z), Omron Platinum BP5450 (HEM-7343T-Z), Walmart Equate Premium 8000 Series UA-8000WM, Walgreens Premium 15+ WGNBPA-960BT, and CVS Series 800 BP3MW1-4YCVS were assessed in an adult general population compared to a mercury sphygmomanometer standard according to the ISO 81060-2:2018/AMD 1:2020 validation procedure. Omron selected the monitors and included three non-Omron monitors because they were from large retail vendors in the United States and these monitors did not have accessible results from validation testing. The BP7450, N = 85, passed both criteria for the standard. Mean (SD) differences in paired SBP and DBP determinations between the test device and reference were 0.5 (7.7) and 2.5 (6.8) mm Hg. The BP5450, N = 86, passed both criteria. Mean (SD) differences in paired SBP and DBP determinations were 1.9 (7.0) and 3.6 (6.4) mm Hg. The UA-8000WM, N = 85, did not meet the first criterion for the standard. Mean (SD) differences in paired SBP and DBP determinations were 2.5 (8.0) and 5.1 (6.4) mm Hg. The WGNBPA-240BT, N = 85, did not meet the first criterion for the standard. Mean (SD) differences in paired SBP and DBP determinations were 7.9 (8.5) and 5.5 (6.7) mm Hg. The BP3MW1-4YCVS, N = 85, did not meet the first criterion for the standard. Mean (SD) differences in paired SBP and DBP determinations were 5.8 (8.7) and 3.1 (5.6) mm Hg. These findings emphasize the importance of verifying the validation status of home blood pressure monitors before use by consumers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。