Electroanalytical Determination of Cysteine Using the Electrodes Based on Ternary Silver-Copper Sulfides

利用基于三元银-铜硫化物的电极进行半胱氨酸的电分析测定

阅读:1

Abstract

The amperometric determination of cysteine, using an electrode based on ternary silver-copper sulfide, is presented. Electrochemical characterizations (using cyclic voltammetry) of three electrode materials revealed that the electrode based on the material that consists of jalpaite (Ag₃CuS₂), doped with a small amount of metallic silver, has the best electrocatalytical performance for cysteine oxidation. For the amperometric determination, 0.142 V at pH 5 and 0.04 V at pH 7 vs. Ag/AgCl, related to the electrocatalytical oxidation of thiol group, were chosen. Electrochemical impedance spectroscopy together with Fourier transform infrared spectroscopy (FTIR) revealed that oxidation takes place on the electrode surface with fouling effect, which does not affect a wide linear working range between 1 μM and 100 μM. Sensitivities, at pH 5 and pH 7, are calculated to be 0.11 μA μM(-1) and 0.10 μA μM(-1), respectively. The detection limits were determined to be 0.036 μM and 0.024 μM for pH 5 and pH 7, respectively. In the presence of uric acid, folic acid, ascorbic acid, and glucose no interference was noticed. This electrode showed remarkable stability and excellent reproducibility. The electrode was exploited for the determination of cysteine in a dietary supplement with the excellent recoveries.