Disposable Pencil Graphite Electrode for Diosmin Voltammetric Analysis

用于地奥司明伏安分析的一次性铅笔状石墨电极

阅读:1

Abstract

Diosmin (DIO) is a naturally occurring flavonoid with multiple beneficial effects on human health. The presence of different hydroxyl groups in diosmin structure enables its electrochemical investigation and quantification. This work presents, for the first time, diosmin voltammetric behavior and quantification on the cost-effective, disposable pencil graphite electrode (PGE). Diosmin oxidation on PGE involves two irreversible steps, generating products with reversible redox behaviors. All electrode processes are pH-dependent and predominantly adsorption-controlled. Differential pulse (DPV) and adsorptive stripping differential pulse (AdSDPV) voltammetric methods have been optimized for diosmin quantification o an H-type PGE, in 0.100 mol/L H(2)SO(4). The linear ranges and limits of detection were for DPV 1.00 × 10(-6)-1.00 × 10(-5) mol/L and 2.76 × 10(-7) mol/L DIO for DPV and 1.00 × 10(-7)-2.50 × 10(-6) mol/L and 7.42 × 10(-8) mol/L DIO for AdSDPV, respectively. The DPV method was successfully applied for diosmin quantification in dietary supplement tablets. The percentage recovery was 99.87 ± 4.88%.