Influence of artefacts generated by titanium and zirconium implants in the study of trabecular bone architecture in cone-beam CT images

钛和锆植入物产生的伪影对锥形束CT图像中小梁骨结构研究的影响

阅读:1

Abstract

OBJECTIVES: To evaluate the influence of artefacts generated by titanium and zirconium implants on trabecular bone architecture assessment through cone-beam CT (CBCT). The influence of kilovoltage (kVp) and metal artefact reduction (MAR) in such analysis was also assessed. METHODS: CBCT scans were obtained with Picasso Trio with or without a titanium or zirconium implants in a posterior region of a mandible using 70, 80 or 90 kVp, with or without MAR. The other acquisition settings were constant: field of view 8 × 5 cm, voxel size 0.2 mm, 5 mA, 24 s and 720 frames. Two volumes of interest (VOIs) were determined comprising trabecular bone mesial and distal to the implant area. The following morphometric parameters were measured: connectivity density (Conn. Dn.), fractal dimension (FD), bone volume fraction (BV/TV), bone surface density (BS/TV), trabecular thickness (Tb. Th.), and trabecular spacing (Tb. Sp.), and compared by multiway ANOVA (α = 0.05). RESULTS: For Conn. Dn. and FD, with MAR, the zirconium group showed values significantly lower than the other groups (p < 0.05). For BV/TV, BS/TV, Tb. Th. and Tb. Sp., the zirconium group showed the highest values, regardless of MAR condition (p < 0.05). MAR increased BS/TV and Tb. Th. values, and decreased FD values for zirconium group. In general, the kVp level did not influence trabecular morphometric parameters. CONCLUSION: The assessment of the trabecular bone architecture was mainly influenced by the expression of the artefacts generated by zirconium implants. MAR decreased the FD and increased the BS/TV and Tb.Th. values regardless of the kVp level.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。