Alveolar bone measurement precision for phosphor-plate images

磷板图像的牙槽骨测量精度

阅读:1

Abstract

OBJECTIVES: The aim was to demonstrate methods for determining measurement precision and to determine the precision of alveolar bone measurements made with a vacuum-coupled positioning device and phosphor plate images. STUDY DESIGN: Subjects were rigidly attached to the x-ray tube by means of a vacuum coupling device and custom cross-arch bite plates. Original and repeat radiographs (taken within minutes of each other) were obtained of the mandibular posterior teeth of 51 subjects, and cementoenamel junction-alveolar crest (CEJ-AC) distances were measured on both sets of images. In addition, x-ray transmission (radiodensity) and AC height differences were determined by subtracting one image from the other. Image subtractions and measurements were performed twice. Based on duplicate measurements, the root mean square standard deviation (precision) and least significant change (LSC) were calculated. LSC is the magnitude of change in a measurement needed to indicate that a true biologic change has occurred. RESULTS: The LSCs were 4% for x-ray transmission, 0.49 mm for CEJ-AC distance, and 0.06 mm for crest height. CONCLUSION: The LSCs for our CEJ-AC and x-ray transmission measurements were similar to what has been previously reported. The LSC for AC height (determined with image subtraction) was <0.1 mm. Compared with findings from earlier studies, this represents a highly precise measurement of AC height. The methods demonstrated for calculating LSC can be used by investigators to determine how large changes in radiographic measurements need to be before the changes can be considered to be (with 95% confidence) true biologic changes and not noise (i.e., equipment/observer error).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。