Space invader: pleural penetration of a hypoglossal nerve stimulator sensor lead

空间入侵者:舌下神经刺激器传感器导线穿透胸膜

阅读:1

Abstract

The mainstay of treatment for obstructive sleep apnea is positive airway pressure therapy, which may be difficult for some patients to tolerate leading to compromised adherence and requiring alternative therapies. Hypoglossal nerve stimulation has become an option for those who meet implantation criteria. Implantation of the device is an ambulatory surgical procedure and is generally well-tolerated, though rare adverse events have been reported. We report an unusual complication of hypoglossal nerve stimulation in a patient who had initial success with this therapy. After 3 years of treatment, the sensor lead penetrated into the pleural space. Components of the hypoglossal nerve stimulation were explanted, and a new sensor lead and generator were reimplanted. The new device was activated, and therapy was successfully resumed. This case demonstrates that there is a potential for a delayed complication of sensor lead penetration into the pleural space, which has only rarely been reported. CITATION: Lou B, Hahn S, Korotun M, Quintero L, Shikowitz M, Greenberg H. Space invader: pleural penetration of a hypoglossal nerve stimulator sensor lead. J Clin Sleep Med. 2021;17(11):2329-2332.