Pullout strength of anterior spinal instrumentation: a product comparison of seven screws in calf vertebral bodies

前路脊柱内固定术的拔出强度:七种螺钉在小牛椎体中的产品比较

阅读:1

Abstract

A lot of new implant devices for spine surgery are coming onto the market, in which vertebral screws play a fundamental role. The new screws developed for surgery of spine deformities have to be compared to established systems. A biomechanical in vitro study was designed to assess the bone-screw interface fixation strength of seven different screws used for correction of scoliosis in spine surgery. The objectives of the current study were twofold: (1) to evaluate the initial strength at the bone-screw interface of newly developed vertebral screws (Universal Spine System II) compared to established systems (product comparison) and (2) to evaluate the influence of screw design, screw diameter, screw length and bone mineral density on pullout strength. Fifty-six calf vertebral bodies were instrumented with seven different screws (USS II anterior 8.0 mm, USS II posterior 6.2 mm, KASS 6.25 mm, USS II anterior 6.2 mm, USS II posterior 5.2 mm, USS 6.0 mm, USS 5.0 mm). Bone mineral density (BMD) was determined by quantitative computed tomography (QCT). Failure in axial pullout was tested using a displacement-controlled universal test machine. USS II anterior 8.0 mm showed higher pullout strength than all other screws. The difference constituted a tendency (P = 0.108) when compared to USS II posterior 6.2 mm (+19%) and was significant in comparison to the other screws (+30 to +55%, P < 0.002). USS II posterior 6.2 mm showed significantly higher pullout strength than USS 5.0 mm (+30%, P = 0.014). The other screws did not differ significantly in pullout strength. Pullout strength correlated significantly with BMD (P = 0.0015) and vertebral body width/screw length (P < 0.001). The newly developed screws for spine surgery (USS II) show higher pullout strength when compared to established systems. Screw design had no significant influence on pullout force in vertebral body screws, but outer diameter of the screw, screw length and BMD are good predictors of pullout resistance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。