Glenohumeral joint injection: a comparative study of ultrasound and fluoroscopically guided techniques before MR arthrography

肩肱关节注射:磁共振关节造影术前超声引导与透视引导技术的比较研究

阅读:1

Abstract

To assess the variability in accuracy of contrast media introduction, leakage, required time and patient discomfort in four different centres, each using a different image-guided glenohumeral injection technique. Each centre included 25 consecutive patients. The ultrasound-guided anterior (USa) and posterior approach (USp), fluoroscopic-guided anterior (FLa) and posterior (FLp) approach were used. Number of injection attempts, effect of contrast leakage on diagnostic quality, and total room, radiologist and procedure times were measured. Pain was documented with a visual analogue scale (VAS) pain score. Access to the joint was achieved in all patients. A successful first attempt significantly occurred more often with US (94%) than with fluoroscopic guidance (72%). Leakage of contrast medium did not cause interpretative difficulties. With US guidance mean room, procedure and radiologist times were significantly shorter (p < 0.001). The USa approach was rated with the lowest pre- and post-injection VAS scores. The four image-guided injection techniques are successful in injection of contrast material into the glenohumeral joint. US-guided injections and especially the anterior approach are significantly less time consuming, more successful on the first attempt, cause less patient discomfort and obviate the need for radiation and iodine contrast.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。