Accuracy of the point-of-care glucose meter for use in calves

用于犊牛的床旁血糖仪的准确性

阅读:1

Abstract

The aim of this study was to evaluate the accuracy and precision of portable blood glucose meters, such as i-STAT 1 and Precision Xceed, for use in calves. Whole blood and plasma samples were obtained from eleven calves that received 2.5 or 5.0% dextrose-containing polyelectrolyte isotonic solutions. Measurements using the i-STAT 1 (r(2)=0.99, P<0.0001) and Precision Xceed (r(2)=0.96, P<0.0001) were well correlated with those by the hexokinase method, which is the gold standard. Although the accuracy of i-STAT 1 was equivalent to that of the hexokinase method, there was an autocorrelation in the residuals between the results from the Precision Xceed and the hexokinase method. Thus, the i-STAT 1 can be used to measure the blood glucose concentration in cattle.