Iodine-based contrast staining improves micro-computed tomography of atherosclerotic coronary arteries

碘基对比染色可改善动脉粥样硬化冠状动脉的微型计算机断层扫描图像。

阅读:1

Abstract

We sought to develop a reversible staining protocol using micro-computed tomography (micro-CT) paired with a radiopaque contrast agent that allows for three-dimensional in situ visualization and characterization of atherosclerotic plaques. Atherosclerotic porcine coronary arteries were dissected from surrounding myocardium and incubated in iohexol at various concentrations and incubation times and then imaged using direct radiography. Line profiles were generated across the artery x-ray to determine effectiveness of the radiopaque contrast agent to penetrate the tissue. Our studies revealed that, to sufficiently delineate tissue constructs, the minimum effective iohexol concentration and incubation time were 240 mgI/mL for 1 hour. Among all groups, 24 hours of de-staining brought radiopacity back to control levels. After iohexol incubation, micro-CT was performed. Our findings demonstrate that extended staining times and a minimum iohexol concentration of 240 mgI/mL are required for effective tissue perfusion, which eliminates the diffusion distribution profile inherent to the ability of the contrast agent to traverse tissue layers.•Iohexol enhances ex vivo micro-CT imaging of atherosclerotic coronary arteries•Iohexol allows for improved tissue segmentation during micro-CT image analysis•Effectiveness of iohexol penetration of the tissue was dependent on concentration and duration of incubation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。