Achieving interlocking nails without using an image intensifier

无需使用图像增强器即可实现互锁钉

阅读:1

Abstract

Interlocking nails are commonly performed using an image intensifier. These are expensive and are not readily available in most resource-poor countries of the world. The aim of this study was to achieve interlocking nailing without the use of an image intensifier. This is a prospective descriptive analysis of 40 consecutive cases seen with shaft fractures of the humerus, femur, and tibia. Fracture fixation was done using Surgical Implant Generation Network (SIGN) nails. Forty limbs in 34 patients were studied. There were 12 females and 22 males, giving a ratio of 1:2. The mean age (years) was 35.75+/-13.16 and the range was 17-70 years. The studied bones were: humerus 10%, femur 65%, and tibia 25%. The fracture lines were: transverse 40%, oblique 15%, and communited 45%. Fracture grades were: closed 90%, grade I, 5%, grade II, 2.5%, and grade IIIA, 2.5%. Surgical approaches were: antegrade 62.5% and retrograde 37.5%. Indications for fixation were: recent fracture 92.5%, non-union 5%, and malunion 3%. Methods of reductions were: open 85% and closed 15%. The mean follow-up period (years) was 1.50+/-0.78. The union time averaged 3 months. Complication was mainly screw loosening due to severe osteoporoses in one case. It is, therefore, concluded that, with the aid of external jigs and slot finders, interlocking can be achieved without an image intensifier.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。