Abstract

BACKGROUND: Implantation of implantable cardioverter-defibrillators (ICD) from the left pectoral region is the standard therapeutical method. Increasing numbers of system revisions due to lead dysfunction and infections will consecutively increase the numbers of right-sided implantations. The reliability of devices implanted on the right pectoral side remains controversially discussed, and the question of testing these devices remains unanswered. METHODS: In a prospectively designed study all 870 patients (60.0 ± 14  years, 689 male) who were treated with a first ICD from July 2005 until May 2012 and tested intraoperatively according to the testing protocol were analyzed. The indication for implantation was primary prophylactic in 71.5%. Underlying diseases included ischemic cardiomyopathy (50%), dilative cardiomyopathy (37%), and others (13%). Mean ejection faction was 27 ± 12%. Implantation site was right in 4.5% and left in 95.5%. RESULTS: Five patients supplied with right-sided ICD (13%, p = 0.02 as compared to left-sided) failed initial intraoperative testing with 21 J. 3 patients were male. The age of the patients failing intraoperative testing with right-sided devices appeared higher than of patients with left-sided devices (p = 0.07). The ejection fraction was 28 ± 8%. All patients reached a sufficient DFT ≤ 21 J after corrective procedures. CONCLUSION: Implantation of ICDs on the right side results in significantly higher failure rate of successful termination of intraoperatively induced ventricular fibrillation. The data of our study suggest the necessity of intraoperative ICD testing in right-sided implanted ICDs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。