Ultrasensitive Electrochemiluminescence Immunoassay Based on Signal Amplification of 0D Au-2D WS(2) Nano-Hybrid Materials

基于0D Au-2D WS(2)纳米杂化材料信号放大的超灵敏电化学发光免疫分析

阅读:1

Abstract

In this study, we proposed a novel Ru(bpy)(3)(2+)-Au-WS(2) nanocomposite (Ru-Au-WS(2) NCs) nano-hybrid electrochemiluminescence (ECL) probe for the highly sensitive detection of carcinoembryonic antigen (CEA). This system utilizes Au nanoparticles (Au NPs) as a bridge to graft the high-performance of a Ru(bpy)(3)(2+) ECL emitter and WS(2) nanosheet with excellent electrochemical performance into an ECL platform, which shows outstanding anodic ECL performance and biosensing platform due to the synergetic effect and biocompatibility of Au NPs and WS(2) nanosheet. Because the ECL intensity of Ru(bpy)(3)(2+) is sensitively affected by the antibody-antigen insulator, a preferable linear dependence was obtained in the concentration range of CEA from 1 pg·mL(-1) to 350 ng·mL(-1) with high selectivity (LOD of 0.3 pg·mL(-1), S/N = 3). Moreover, the ECL platform had good reproducibility and stability and exhibited excellent anti-interference performance in the detection process of CEA. We believe that the platform we have developed can expand the opportunities for the detection of additional high specificity-related antibodies/antigens and demonstrate broad prospects for disease diagnosis and biochemical research.