Comparison of the corrected intraocular pressure by tonopachy with that of Goldmann applanation tonometry in normal and glaucomatous patients

比较正常人和青光眼患者使用眼压计测量的校正眼压与使用Goldmann压平式眼压计测量的校正眼压

阅读:1

Abstract

PURPOSE: To compare corrected intraocular pressure (IOP) by tonopachy with that of Goldmann applanation tonometry (GAT) in normal and glaucomatous patients. METHODS: In this cross-sectional study, IOP and central corneal thickness (CCT) were measured in 426 eyes (213 normal eyes and 213 glaucomatous eyes) of 426 patients by tonopachy followed by GAT and ultrasound pachymetry. IOP was corrected for CCT by in-built formula in tonopachy and Ehlers correction factor for Goldmann tonometer. Limits of agreements were assessed using Bland-Altman plots. Intraclass correlation coefficient was calculated to estimate the absolute agreement between single and average measurements of IOP and CCT of tonopachy with that of Goldmann tonometer and ultrasound pachymetry respectively. RESULTS: Mean corrected IOP measured with tonopachy and GAT in glaucomatous eyes was 17.63 ± 5.04 mmHg and 19.42 ± 5.83 mmHg, and in controls it was 13.4 ± 2.5 mmHg and 16.2 ± 3.1 mmHg, respectively. Limits of agreement ranged from -4.63 to +9.25 mmHg for total population (mean = 2.31), -6.01 to +9.59 mmHg (mean = 1.79) for glaucoma group and -2.99 to +8.65 mmHg (mean = 2.83) for controls. Intraclass correlation coefficient for IOP measurement between tonopachy and Goldmann tonometer was 0.84 for total population, 0.85 for glaucoma group, and 0.63 for controls, respectively. CONCLUSION: Corrected IOP obtained by tonopachy showed moderate agreement with GAT and it is more in glaucoma patients than controls. Thus, tonopachy can be used as a screening tool, but cannot replace GAT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。