Enumeration of lymphocyte subsets using flow cytometry: Effect of storage before and after staining in a developing country setting

利用流式细胞术对淋巴细胞亚群进行计数:发展中国家环境下染色前后储存的影响

阅读:1

Abstract

Lymphocyte subset estimations by flow cytometry in population-based studies require transportation of samples from the field site to the laboratory. As samples arrive late in the day they have to wait overnight before being processed. The effect of two possible approaches, sample storage for 24 h before staining and immediate staining with analysis after 24 h and 48 h were evaluated. Two sets of experiments were performed with EDTA (ethylenediamine tetra-acetate) anticoagulated peripheral blood. In the first experiment, after collection, each sample was divided into two portions. One portion was stained at the time of blood collection and the other 24 h later after keeping it at room temperature (38-45°C). In the second experiment, blood samples were stained within 1-2 h. Each sample was analyzed immediately upon completion of staining process and subsequently after 24 h and 48 h of storage at 4°C. Results suggest that blood collected in EDTA can be processed using whole blood lysis method, after storage at room temperature (38-45°C) for 24 h with some but not significant alteration in T-cell subsets. Storage at 4°C after staining for 24 h results in a lesser and insignificant loss of cells or alteration of T-cell subsets and may be the method of choice.