Abstract

INTRODUCTION: Carotid angioplasty and stenting (CAS) has been demonstrated to be safe and an acceptable alternative to surgery. Stent malpositioning can occur during the maneuvers of delivering; technical errors can lead to proximal or distal slipping of the stent that needs the placement of additional pieces. PRESENTATION OF CASE: We describe the case of a postoperative dislocation of a carotid stent that happened 1 year after placement. After the first ultrasound control confirmed the correct position of the Stent the following one, executed 9 months later, showed a severe restenosis due to a proximal dislocation of the stent. The problem was solved with the positioning of a further one more distally. DISCUSSION: We observe the possibility of Carotid Wallstent shortening during the implant due to an incorrect placement or sizing, but a delayed displacement is a rare complication that, we highlight, can occur after CAS and that needs to be considered at the moment of the preoperative planning. CONCLUSION: After CAS a closed ultrasound follow up is advisable for a long time in order to detect unexpected complications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。