Conclusions
It was found possible to complement a commercially available kit for quantitative serum steroid profiling based on isotope dilution LC-MS/MS by implementing automated on-line SPE, thereby improving the practicality and robustness of the measurement procedure.
Methods
Sample preparation was performed using automated on-line SPE on a high-pressure stable extraction column. Chromatographic separation, including isobaric compounds, was achieved using a 0.25 mM ammonium fluoride-methanol gradient on a small particle size biphenyl column. Standard compounds and internal standard mixtures of two panels of a commercially available kit were combined to achieve an optimized and straightforward detection of 15 endogenous steroids. Validation was performed according to the European Medicines Agency (EMA) guidelines with slight modifications.
Results
Validation was successfully performed for all steroids over a clinically relevant calibration range. Deviations of intra- and inter-assay accuracy and precision results passed the criteria and no relevant matrix effects were detected due to highly effective sample preparation. External quality assessment samples showed the applicability as a routine diagnostic method, which was affirmed by the analyses of anonymized clinical samples. Conclusions: It was found possible to complement a commercially available kit for quantitative serum steroid profiling based on isotope dilution LC-MS/MS by implementing automated on-line SPE, thereby improving the practicality and robustness of the measurement procedure.