Design, Fabrication and Testing of an Insertable Double-Imaging-Region Gradient Coil

可插入式双成像区梯度线圈的设计、制造和测试

阅读:1

Abstract

We have constructed a small-bore insertable gradient coil with two linear gradient imaging regions and interfaced it with an MRI scanner. We have also constructed an RF system capable of transmitting or receiving in both regions simultaneously.Designs for conductor placement for two-region X-, Y- and Z-gradient coils were optimized by simulated annealing. Wire patterns for each axis were chosen that gave low inductance, reasonable homogeneity over a large imaging volume and high efficiency (gradient field per-unit-current).Imaging was performed on a Siemens 3T TIM Trio scanner equipped with three additional gradient amplifier channels and a second RF/gradient array controller. Phantoms were placed in the two imaging regions as well as the central non-imaging region to test gradient homogeneity and crosstalk between regions. Images acquired simultaneously in the two regions showed very little signal crosstalk between imaging regions and even less signal from the central, non-imaging region.When combined with an overlapping single-region gradient insert, extended field-of-view (FOV) imaging will be possible without moving the table or the subject and without increasing nerve stimulation. Construction and testing of a two-region gradient coil insert is a necessary intermediate step as a proof of concept for an extended field of view, contiguous, three-region human-sized gradient system.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。