Automatic Coregistration Between Coronary Angiography and Intravascular Optical Coherence Tomography: Feasibility and Accuracy

冠状动脉造影与血管内光学相干断层扫描的自动配准:可行性和准确性

阅读:2

Abstract

This study sought to evaluate a novel approach for automatic coregistration of optical coherence tomography (OCT) and coronary angiography. Lumen diameters and side branches from both coronary angiography and OCT were used to create 2 feature sets. Subsequently, a 2-step coregistration approach was performed on the feature sets for matching of each OCT cross section on the angiographic centerline. For validation, all side branches with ≥1.0 mm diameter were identified and used as paired fiduciary landmarks. Geographical error was defined as the distance between the automatically coregistered and the true-paired landmarks. Altogether 212 vessels from 181 patients were analyzed. Mismatch of coronary angiography and OCT occurred in 64 of 1,530 reference landmarks. Median geographical error was 0.32 (interquartile range: 0.00-0.56) mm. The mean time for coregistration was 20.69 ± 1.07 seconds. In conclusion, fast and automatic coregistration of OCT and angiography using a single standard angiographic loop is feasible and accurate.