Evaluation of a contact lens-embedded sensor for intraocular pressure measurement

评估一种用于眼内压测量的隐形眼镜嵌入式传感器

阅读:1

Abstract

PURPOSE: To evaluate a novel contact lens-embedded pressure sensor for continuous measurement of intraocular pressure (IOP). METHODS: Repeated measurements of IOP and ocular pulse amplitude (OPA) were recorded in 12 eyes of 12 subjects in sitting and supine positions using 3 configurations of the dynamic contour tonometer: slit-lamp mounted (DCT), hand-held (HH), and contact lens-embedded sensor (CL). The IOP and OPA for each condition were compared using repeated measures ANOVA and the 95% limits of agreement were calculated. RESULTS: The sitting IOP (mean and 95% CI) for each configuration was DCT: 16.3 mm Hg (15.6 to 17.1 mm Hg), HH: 16.6 mm Hg (15.6 to 17.6 mm Hg), and CL: 15.7 mm Hg (15 to 16.3 mm Hg). The sitting OPA for each configuration was DCT: 2.4 mm Hg (2.1 to 2.6 mm Hg), HH: 2.4 mm Hg (2.1 to 2.7 mm Hg), and CL: 2.1 mm Hg (1.8 to 2.3 mm Hg). Supine IOP and OPA measurements with the CL and HH sensors were both greater than their corresponding sitting measurements, but were significantly less with the CL sensor than the HH sensor. The mean difference and 95% Limits of Agreement were smallest for the DCT and CL sensor comparisons (0.7+/-3.9 mm Hg) and widest for the CL and HH sensors (-1.9+/-7.25 mm Hg); these wider limits were attributed to greater HH measurement variability. CONCLUSIONS: The CL sensor was comparable to HH and DCT sensors with sitting subjects and is a viable method for measuring IOP and OPA. Supine measurements of IOP and OPA were greater than sitting conditions and were comparatively lower with the CL sensor. HH measurements were more variable than CL measurements and this influenced the Limits of Agreement for both sitting and supine conditions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。