Comparison of Laser-Sintered and Milled Patient-Specific Reconstruction Plates for Complications and Outcomes in Mandibular Defects-Comparative Analysis of a Single-Center Cohort

激光烧结和铣削患者特异性重建钢板在下颌骨缺损并发症和疗效方面的比较——单中心队列研究的比较分析

阅读:1

Abstract

CAD/CAM-manufactured implants are increasingly becoming the standard in current therapy. The question of whether the manufacturing-related rougher surface of selective laser fusion plates compared to milled, smoother reconstruction plates leads to increased postoperative complications such as infections, plate exposure, and fistulas has not yet been determined. A retrospective analysis of 98 patients who underwent surgical treatment with either a selective laser fusion plate or a milled reconstruction plate at our hospital was performed. The only significant predictors of the revision risk were the operation time and use of antiresorptive medication. In the KLS Martin(®) group, the risk of revision decreased by approximately 20% for each additional hour by which the operation time was increased (OR = 0.81). In the Depuy Synthes(®) group, the risk of revision increased by approximately 11% with each additional hour of operative time (OR = 0.81 × 1.37 = 1.11). Both groups showed no significant differences in the number of necessary revision surgeries as well as inpatient complications. In summary, we can say that the assumption that additively manufactured reconstruction plates have a rougher surface due to selective laser melting and thus make plaque accumulation and revisions more likely has not been confirmed. Overall, it seems imperative to select further studies regarding the clinical outcome depending on the selected plate system.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。