Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

手术扭矩限制器的有效性:比较钻孔和手工螺钉插入锁定钢板的模型

阅读:1

Abstract

BACKGROUND: The objective of this study is to analyse the effectiveness of the surgical torque limiter during operative use. The study also investigates the potential differences in torque between hand and drill-based screw insertion into locking plates using a standardised torque limiter. METHODS: Torque for both hand and power screw insertion was measured through a load cell, registering 6.66 points per second. This was performed in a controlled environment using synthetic bone, a locking plate and locking screws to simulate plate fixation. Screws were inserted by hand and by drill with torque values measured. RESULTS: The surgical torque limiter (1.5 Nm) was effective as the highest recorded reading in the study was 1.409 Nm. Comparatively, there is a statistically significant difference between screw insertion methods. Torque produced for manually driven screw insertion into locking plates was 1.289 Nm (95 % CI 1.269-1.308) with drill-powered screw insertion at 0.740 Nm (95 % CI 0.723-0.757). CONCLUSIONS: The surgical torque limiter proved to be effective as per product specifications. Screws inserted under power produce significantly less torque when compared to manual insertion by hand. This is likely related to the mechanism of the torque limiter when being used at higher speeds for which it was designed. We conclude that screws may be inserted using power to the plate with the addition of a torque limiter. It is recommended that all screws inserted by drill be hand tightened to achieve adequate torque values.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。