Exploring the Feasibility of Estimating Intraocular Pressure Using Vibrational Response of the Eye: A Methodological Approach

探索利用眼部振动响应估算眼内压的可行性:一种方法学方法

阅读:1

Abstract

This study addresses the limitations of current tonometry techniques by exploring vibroacoustic properties for estimating intraocular pressure (IOP), a key diagnostic parameter for monitoring glaucoma-a significant risk factor for vision loss. Utilizing vivo porcine eyeballs, we investigated the relationship between IOP and the nonlinear vibration transfer function ratio (NVTFR). Through applying varying vibration levels and analyzing responses with transfer function analysis and univariate regression, we identified a strong negative correlation between NVTFR and IOP, evidenced by a Pearson correlation coefficient of -0.8111 and significant results from generalized linear model (GLM) regression (p-value < 0.001). These findings indicate the potential of NVTFR as a vital indicator of IOP changes. Our study highlights the feasibility of using vibroacoustic properties, specifically NVTFR, to measure IOP. While further refinement is necessary for in vivo application, this approach opens new possibilities for non-invasive and patient-friendly IOP monitoring, potentially enhancing ophthalmology diagnostic techniques and providing a foundation for future research and development in this critical area.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。