Validation and reproducibility assessment of tigecycline MIC determinations by Etest

采用Etest法测定替加环素MIC的验证和重复性评估

阅读:1

Abstract

A multicenter study was conducted to validate Etest tigecycline compared to the Clinical Laboratory Standards Institute reference broth microdilution and agar dilution methodologies. A large collection of gram-negative (n = 266) and gram-positive (n = 162) aerobic bacteria, a collection of anaerobes (n = 385), and selected collections of nonpneumococcal streptococci (n = 369), Streptococcus pneumoniae (n = 372), and Haemophilus influenzae (n = 372) were tested. Strains with reduced susceptibility to tigecycline were used with all test methods. The Etest showed excellent inter- and intralaboratory reproducibility for all organism groups tested regardless of the test methodology. The essential agreement values with the reference method (+/-1 dilution) were >99% for the collection of gram-negative and gram-positive aerobes; >98% for the S. pneumoniae, H. influenzae, and anaerobe collections; and 100% for the group of nonpneumococcal streptococci. These results validate the performance accuracy and utility of Etest tigecycline and verify the reproducibility of this convenient predefined gradient methodology for tigecycline susceptibility determination.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。