Clinical comparison of Sentinel, a novel blood culture system, with radiometric Bactec 460 and Isolator 10 in the detection of streptococcal and anaerobic bacteraemias

新型血培养系统 Sentinel 与放射性血培养系统 Bactec 460 和 Isolator 10 在检测链球菌和厌氧菌血症方面的临床比较

阅读:2

Abstract

AIMS: To compare the performance of the Sentinel blood culture system with two other systems for the recovery of streptococci and anaerobes. METHODS: Blood cultures were taken from 55 patients one to two minutes after dental extraction. The samples were tested by the radiometric Bactec 460; the Isolator 10, which works by lysis centrifugation; and Sentinel, a fully automated system, which detects bacterial growth by changes in the voltage between two electrodes in the media. Positive samples were subcultured and streptococci and anaerobes were further identified. Terminal subcultures were performed on all negative samples. RESULTS: Sentinel was comparable with Bactec, with Sentinel recovering 20 streptococci and 14 anaerobes; Bactec isolated 26 streptococci and 15 anaerobes. The recovery of Streptococcus sanguis was significantly better from Bactec than Sentinel. The Isolator 10 recovered significantly fewer streptococci and anaerobes than either Bactec or Sentinel. Sentinel was noticeably quicker in detecting anaerobes than Bactec 460. However, there was no comparable difference with streptococci. Contaminants were recovered from 10% of Isolator 10, 2.7% of Bactec, and 7.2% of Sentinel bottles. CONCLUSIONS: Sentinel and Bactec 460 were broadly comparable in the detection of viridans streptococci and oral anaerobes, and both systems were significantly better than the Isolator 10. However, the prototype Sentinel, was significantly poorer than Bactec for the recovery of S sanguis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。