Contamination of cultures processed with the isolator lysis-centrifugation blood culture tube

使用分离器裂解离心血培养管处理的培养物受到污染

阅读:1

Abstract

Overall contamination (on- plus off-streak) of the Isolator (Du Pont Co.) blood culture tube (23%) was greater than that of a conventional broth blood culture bottle (0.6%) or that of a biphasic blood culture bottle (1.3%). To determine the source of this contamination, Isolator cultures of blood from 59 healthy volunteers and of sterile broth from 60 vials were made. A total of 37% of the blood cultures and 22% of the broth cultures were contaminated (P = 0.06). Staphylococcus epidermidis-contaminated cultures represented 31 and 10% of the blood and broth cultures, respectively (P = 0.06). Contamination of plates processed on a bench top, in front of horizontal laminar flow, and in a biological safety cabinet with vertical laminar flow were compared. Processing plates in a biological safety cabinet resulted in a significant reduction in the number of contaminated plates (P less than 0.05). The contamination rate for 7,874 Isolator blood cultures processed in the biological safety cabinet was significantly decreased to 6.7% on-streak (9.3% on- plus off-streak). Contamination of Isolator-processed blood cultures originated from the laboratory and the patient. The former can be reduced by inoculating plates in a vertical laminar flow biological safety cabinet and by maintaining adequate quality control of media. The latter may be unavoidable.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。